Printing or Sending Results

At the Results screen, you can send the QC analysis results to your printer or to a data
management system or an LIS.
To print a report, select the
To send the report to a data management system or an LIS, select
The report identifies the control and lists the target ranges as you defined them in Setup. A
legend at the bottom explains the results symbols used in the report.
If you select RiliBÄK range checking in Setup, the report contains an additional line with
the parameter target value and the percentage difference between the measured value and
target. The report also displays a "Failed" message if the measured value exceeds the
RiliBÄK range. Refer to Setting Up QC Options‚ page 8-2.

Result Flags

The system displays the following symbols in red on the results screen and in the printed
report:
This symbol...
¦
Ø
-----↑
-----↓
-----?
02087462 Rev. V
Rapidlab 1200 Operator's Guide: Quality Control
button.
Print
Indicates that...
the result is above the target range.
the result is below the target range.
the result is above the reporting range.
the result is below the reporting range.
the system has an atypical response when measuring this
parameter.
.