Flushing the Upper Connector

1. Connect the syringe tubing to the upper connector for the wash reagent.
1
2
Figure 5-26
2. Flush with a minimum of 3 mL water.
3. Flush with a minimum of 12 mL air.
4. Collect the water in an absorbent swab or equivalent.
02087462 Rev. V
Water collection area
Upper connector
Reagent Manifold