Setting the maximum filament inverter frequency

The maximum filament inverter freqency is 25 kHz and can be set with P3 on D915.
Proceed as follows:
Switch POLYMOBIL Plus OFF.
Remove fuse F6 on D925.
Connect the oscilloscope to D915 at TP "CAL" and "GND".
Switch POLYMOBIL Plus ON (Err 03 is blinking).
The oscillogram must appear like the one in Fig. 12.
CH1: D915 test point "CAL"
20 µs /Div
Fig. 12
Setting the maximum filament inverter frequency with potentiometer P3 on D915.
POLYMOBIL Plus

SPR8-125.061.02

Rev. 01
01.00
Measurements
Page 12 of 20