Shutting Down The System - Siemens Rapidlab 1200 Operator's Manual

Table of Contents

Advertisement

2-6

Shutting Down the System

Use this procedure to remove power from the system for 24 hours or less.
CAUTION:
cartridges are installed. Cartridges installed in the system remain stable for 6
hours without power.
CAUTION:
this procedure to shut down the system.
1. Select
2. When prompted, turn the power switch off.
The power switch is on the back panel of the system.
WARNING:
mixture with air, O
explosive environment. Refer to Protecting Yourself from Electrical Hazards‚
page A-3 for recommended precautions when working around electricity.
3. To restore power to the system, turn the power switch on.
02087462 Rev. V
Rapidlab 1200 Operator's Guide: Operating the System
Do not remove power from the system for more than 6 hours if
To prevent damage to the hard drive and permanent loss of data, use
Status > Shutdown > Yes
Do not operate the system in the presence of flammable anesthetic
, or nitrous oxide. The risk of explosion exists in a potentially
2
.