Analyzing Samples; Analyzing Syringe Samples - Siemens Rapidlab 1200 Operator's Manual

Table of Contents

Advertisement

Rapidlab 1200 Operator's Guide: Operating the System

Analyzing Samples

This section contains information about analyzing the following samples:
Syringe samples
Capillary samples
Microsamples
pH and pH, Glu, and Lac samples
tHb samples

Analyzing Syringe Samples

Use this procedure to analyze patient blood samples using a syringe.
BIOHAZARD:
Refer to Appendix A, Protecting Yourself from Biohazards for recommended
precautions when working with biohazardous materials.
1. Observe all rules and guidelines in Collecting Patient Samples‚ page 2-7.
NOTE:
that the system is busy, select
available, wait until the message is not displayed to analyze the patient
sample.
2. At the Analysis screen, select the patient sample type (arterial, venous, mixed
venous).
The default type is a syringe of arterial blood.
3. Scan the barcode for the patient sample, if required.
4. Insert the syringe into the sample port.
5. Select
Wear personal protective equipment. Use universal precautions.
If you have a priority sample, but a message is displayed indicating
.
Analyze
to interrupt the system. If Cancel is not
Cancel
2-11
02087462 Rev. V

Advertisement

Table of Contents
loading