Rapidlab 1200 Operator's Guide: System Configuration
NOTE:
7. Select
The default panel is selected automatically. Parameters that are not in the
default panel do not display. You can deselect the default panel at the Analysis
screen, then select a different custom panel or select other parameters.
Microsample panels display a µ character at the top-left of the panel.

Setting up Sample Types

You can select the default patient sample type that is selected at the Analysis
screen. The default sample type, which is either arterial syringe or capillary tube,
is displayed with a check mark at the Analysis screen. Operators can select
Continue
another sample type.
1. At the Setup screen, select
2. Select the sample type button that you want as the default selection.
3. Select

Setting up Parameters

When you select Parameters at the Setup Window, the following options are
available:
Parameter On/Off
Parameter Units
Demographic Units
Values

Using the Parameters On/Off Screen

At the Setup screen, select
parameters you want the system to report. Use the following guidelines when
using the Parameters On/Off screen:
The system displays certain parameters as gray on the Parameters On/Off
screen and you cannot select them.
These parameters are either not available on the system or other required
parameters were not selected in Setup. For example on Rapidlab1240 and
1260 systems, tHb is not available, and on Rapidlab 1245 and 1265 systems,
if tHb is not selected other parameters that require tHb are not available.
Only the first panel in the first set can be selected as a default panel.
Save > Return > Analysis
to begin analysis of the default patient sample type, or they can select
Sample > Sample Type
.
Save
Parameters > Parameters On/Off
.
.
8-13
to select the
02087462 Rev. V