Replacing The Co-Ox Module Tubing - Siemens Rapidlab 1200 Operator's Manual

Table of Contents

Advertisement

Rapidlab 1200 Operator's Guide: Maintenance
b. Connect the new tubing to the pressure sensor bubbler and to the reagent
6. Dispose of the tubing according to your institution's protocol.
7. Close the reagent cartridge door and then close the front cover.
8. If you have a Rapidlab 1245 or 1265 system and are replacing the CO-ox
tubing, continue with step 2 of Replacing the CO-ox Module Tubing‚
page 5-13.
9. Close the measurement module door.
10. Select
11. Select

Replacing the CO-ox Module Tubing

You need a CO-ox tubing kit for this task.
BIOHAZARD:
Refer to Appendix A, Protecting Yourself from Biohazards for recommended
precautions when working with biohazardous materials.
1. Select
2. Select
3. Lift the door to the measurement module.
Refer to Preparing for Maintenance Procedures‚ page 5-1.
manifold.
.
No
Return > Analyze > Yes
Wear personal protective equipment. Use universal precautions.
Status > Diagnostics > Cartridges
and remove the wash cartridge.
Eject W Cartridge
.
.
5-13
02087462 Rev. V

Advertisement

Table of Contents
loading

Table of Contents