Replacing The System Fuses - Siemens Rapidlab 1200 Operator's Manual

Table of Contents

Advertisement

Rapidlab 1200 Operator's Guide: Maintenance
4. Press the tubing around the outside of the rollers.
5. Place the upper tubing cuff onto the upper side of the platen.
6. Turn the roller cage clockwise to gently work the new tubing between the
platen and the roller cage.
7. Connect the CO-ox pump tubing to the reagent manifold tubing.
8. Close the front cover.
9. Select
10. Perform a full calibration.
For information about performing calibrations, refer to Recalling Calibration
Results‚ page 3-4.

Replacing the System Fuses

Replace both fuses if 1 or both of the fuses are blown.
Before replacing the fuses, verify that power is being supplied to the system:
The power switch on the back panel is turned on.
The power cord is firmly connected to the system and to the electrical outlet.
The electrical outlet is working.
You need the following materials for this task:
Two fuses of the appropriate rating
Screwdriver
The Rapidlab 1200 systems use the following fuses for the voltages shown:
Voltage
100–240V
WARNING:
damage to the system, remove power from the system as described in this
procedure.
CAUTION:
cartridges are installed. Cartridges installed in the system remain stable for 6
hours without power.
.
Return
Fuse Rating
1.6A Slo Blo
Do not proceed while the power is on. To prevent electrical shock or
Do not remove power from the system for more than 6 hours if
Fuse Type
5 x 20 mm
02087462 Rev. V
5-47