02087462 Rev.V
Rapidlab 1240 Recovery and Precision Testing—pCO
n
WRSD
Micro-
24
1.91
syringe
Micro-
24
2.97
syringe
Micro-
24
6.74
syringe
Rapidlab 1240 Recovery and Precision Testing—pO
n
WRSD
Capillary
24
0.34
Capillary
24
0.35
Capillary
24
0.89
Capillary
24
1.47
Capillary
24
0.79
Capillary
24
4.02
Capillary
24
1.53
Capillary
24
4.87
Capillary
24
10.28
Syringe
45
0.36
Syringe
48
0.42
Syringe
48
0.6
Syringe
48
1.15
Syringe
48
1.13
Syringe
45
2.26
Syringe
47
3.81
Syringe
48
4.69
Syringe
48
9.97
Micro-
12
0.21
capillary
Micro-
12
0.35
capillary
Micro-
12
0.7
capillary
Micro-
12
0.64
capillary
Rapidlab 1200 Operator's Guide: System Specifications
*
Observed
Expected
101.7
100.2
138.3
136
241.4
229
*
Observed
Expected
27.9
28.6
52
50
87.7
85.5
101.4
107.1
153.7
149.6
198
199.9
254.1
249.4
374.8
378.4
630
649.7
28.1
28.6
52.5
50.1
88.2
84.9
103.2
107.3
153.7
148.6
196.2
200.3
253.9
247.7
379.5
379.2
636.6
651.1
30.7
28.5
54
49.9
95.1
85.6
106.7
107
2
Recovery
WRCV
101.5
1.9
101.7
2.1
105.4
2.8
2
Recovery
WRCV
97.7
1.2
104
0.7
102.5
1
94.7
1.5
102.7
0.5
99
2
101.9
0.6
99
1.3
97
1.6
98
1.3
104.9
0.8
103.9
0.7
96.2
1.1
103.4
0.7
98
1.2
102.5
1.5
100.1
1.2
97.8
1.6
107.5
0.7
108
0.6
111
0.7
99.7
0.6