Siemens Rapidlab 1200 Operator's Manual page 416

Table of Contents

Advertisement

E-56
Table E-55
Level Mode
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
*
WRSD = within-run standard deviation
02087462 Rev.V
Rapidlab 1265 Recovery and Precision Testing—pCO
N
WRSD
Micro-
12
0.22
capillary
Micro-
12
0.28
capillary
Micro-
12
0.24
capillary
Micro-
12
0.18
capillary
Micro-
11
0.36
capillary
Micro-
12
0.53
capillary
Micro-
12
1.69
capillary
Micro-
12
3.48
capillary
Micro-
12
3.73
capillary
Micro-
24
0.21
syringe
Micro-
24
0.64
syringe
Micro-
24
0.45
syringe
Micro-
24
1.11
syringe
Micro-
24
0.96
syringe
Micro-
24
1.84
syringe
Micro-
24
4.35
syringe
Micro-
24
1.42
syringe
Micro-
24
5.21
syringe
Rapidlab 1200 Operator's Guide: System Specifications
*
Observed
Expected
6.8
7.1
21.3
21.3
28.1
28.4
35.9
35.8
48.9
49.8
70.2
71.1
96.4
100.2
133.2
135.1
221.7
229.1
6.8
7.1
21.5
21.4
28.3
28.6
34.3
35.7
47.6
49.5
71.1
71.4
95.6
99.8
130.9
134.3
217.7
228.2
2
Recovery
WRCV
95.2
3.2
99.7
1.3
99.0
0.9
100.3
0.5
98.3
0.7
98.9
0.8
96.2
1.7
98.6
2.6
96.8
1.7
96.8
3.1
100.4
3.0
99.1
1.6
96.2
3.2
96.3
2.0
99.5
2.6
95.7
4.6
97.4
1.1
95.4
2.4

Advertisement

Table of Contents
loading

Table of Contents