Performing The Wavelength Cal Test; Performing The Co-Ox Sample Chamber Test - Siemens Rapidlab 1200 Operator's Manual

Table of Contents

Advertisement

6-46

Performing the Wavelength Cal Test

The Wavelength Cal test verifies the proper operation of the CO-ox neon lamp and prints
the results providing the system ID, the date and time of the report, and as represented in
Table 6-7.
To perform the Wavelength Cal Report, select
Table 6-7
Row number
1
2
3
4
5
6
7

Performing the CO-ox Sample Chamber Test

This test is only performed on Rapidlab 1245 and 1265 Systems.
Use this procedure to test the ability of the CO-ox sample chamber to move to the open
and close positions. The systems displays the pass or fail results on a printed report.
1. Select
2. Enter your password, if necessary.
3. Select
The sample chamber moves to the open position.
4. Select
5. Select
The system prints the test results Passed or Failed.
6. Wait while the system returns the sample chamber to the correct position.
7. Select
8. To return to the Status screen, select
02087462 Rev. V
Wavelength Cal Report
Data
Mean
Standard deviation
Measured position
Baseline position
Shift
Integration time from last zero
Serial number
.
Status > Diagnostics
.
CO-ox > Start
to move the sample chamber to the closed position.
Change
.
Stop
.
Return
Rapidlab 1200 Operator's Guide: Troubleshooting
CO-ox > Wavelength Cal
.
Return
.

Advertisement