Rapidlab 1200 Operator's Guide: Troubleshooting
2. To remove obstructions from the sample tubing in the reagent manifold, push
the clot-removal line through the fluid connector on the reagent manifold.
1
2
Figure 6-9
3. Remove obstructions from the preheater:
a. Remove the pO
Use gauze to catch waste material cleared by the clot-removal
NOTE:
line.
Sample connector
Fluid connector
Sample Tubing in the Reagent Manifold
and pCO
2
(gas) sensors.
2
6-25
02087462 Rev. V