Siemens Rapidlab 1200 Operator's Manual page 325

Table of Contents

Advertisement

Rapidlab 1200 Operator's Guide: System Configuration
NOTE:
IEC standards (i.e. IEC 950 for data processing equipment and IEC 60601-1 for
medical equipment). All configurations should comply with system standard IEC
60601-1. When you connect additional equipment to the signal input or output
port, the system must comply with the requirements of IEC 60601-1. Consult your
local technical support provider or distributor if necessary.
WARNING:
damage to the system, remove power from the system as described in this
procedure.
1. Select
CAUTION:
cartridges are installed. Cartridges installed in the system remain stable for 6
hours without power.
2. At the Shutdown screen, turn the power switch off.
3. Connect the interface cable to the Rapidlab 1200 system and to the LIS:
a. Connect the 9-pin connector on the cable to the serial (RS-232) port on
b. Connect the other connector on the cable to the LIS.
4. Turn the power switch on.
5. Select the communication settings at the Rapidlab 1200 system:
a. Select
b. If required, enter your password.
c. Select
d. Select
e. Select the communication settings that the LIS requires:
Setting
Baud
Parity
Data Bits
f.
6. At the Communications screen, select
If an error message displays, refer to Viewing System Messages‚ page 6-74.
7. Select
8. Complete any setup at the LIS that may be required to communicate with the
Rapidlab 1200 system.
External devices connected to the digital interfaces must be certified to
Do not proceed while the power is on. To prevent electrical shock or
Status > Shutdown
Do not remove power from the system for more than 6 hours if
the back panel of the Rapidlab 1200 system.
.
Status > Setup
Printer and Devices > Communications
RS-232 > Configure
Default
9600
Even
8
Select
.
Save
.
Return
.
.
Options
1200, 2400, 4800, 9600, 19200
None, Odd, Even
7, 8
.
Save
8-31
.
02087462 Rev. V