6-18
2. Replace the pO
a. Push the black spring-loaded latch assembly (item 1, Figure 6-4), located at the
1
2
3
Figure 6-4
b. Remove the pO
02087462 Rev. V
sensor with the clot clearing tool.
2
right of the 1200 Reference sensor, to the right, until the mechanism clicks in
place.
spring-loaded latch assembly
top of the pO
sensor
2
lip of pO
sensor
2
Measurement Module Compartment
sensor by lifting the sensor lip (item 2, Figure 6-4) up at the
2
bottom while pressing the top of the sensor (item 3, Figure 6-4) inward.
Rapidlab 1200 Operator's Guide: Troubleshooting