Rapidlab 1200 Operator's Guide: Troubleshooting
b. If using a standard syringe, hold the syringe firmly.
7. Press the syringe plunger gently into the clot clearing tool, while holding
gauze against the sample port of the reagent cartridge, to catch the obstruction
and any excess liquid. The sample port location is shown in Figure 6-8.
1
Figure 6-8
BIOHAZARD:
practices of your workplace.
8. Remove the syringe, and discard the gauze.
If an obstruction is cleared, it is expelled, along with any excess liquid,
through the sample port on the reagent cartridge.
9. Reinstall the pO
a. Push the black spring enclosure assembly to the right, until the
b. Remove the syringe.
c. Remove the clot clearing tool, by lifting the clot clearing tool lip up at the
d. Reinstall the pO
Sample port of reagent cartridge in position 2
Sample Port of Reagent Cartridge
Discard the gauze according to the established biohazard disposal
sensor.
2
mechanism clicks in place.
bottom while pressing the top of the tool inward.
sensor.
2
6-23
02087462 Rev. V