Flushing The Reagent Manifold Tubing - Siemens Rapidlab 1200 Operator's Manual

Table of Contents

Advertisement

5-58

Flushing the Reagent Manifold Tubing

This section applies to the Rapidlab 1245 and 1265 systems.
1. Disconnect the reagent manifold tubing from the CO-ox pump tubing.
2. Connect the syringe to the waste tubing.
1
2
3
Figure 5-32
3. Flush the waste tubing with a minimum of 3 mL water.
4. Flush the waste tubing with a minimum of 12 mL air.
5. Collect the water in an absorbent swab or equivalent.
6. Reconnect the reagent manifold tubing from the CO-ox pump tubing.
02087462 Rev. V
Connector
Waste tubing
Water collection area
Reagent Manifold–Rapidlab 1245 and 1265 Systems
Rapidlab 1200 Operator's Guide: Maintenance