1. Select
The system displays the Shutdown screen and stops all fluidic sequences so that no
damage to the cartridges can occur.
2. Turn the power off and disconnect the power cord from the electrical outlet.
3. Open the fuse holder:
a. Insert a small flat-head screwdriver in the tabs on each side of the fuse holder.
b. Pull the fuse holder out from the compartment as far as possible.
1
Figure 5-24
4. Remove the old fuses and dispose of them according to your institution's protocol.
5. Install new fuses.
6. Slide the fuse holder into the fuse compartment.
7. Restore power:
a. Reconnect the power cord to the system.
b. Reconnect the power cord to the electrical outlet.
c. Turn the power on.
02087462 Rev. V
Status > Shutdown > Yes
Fuse holder
Fuse Holder