Siemens Rapidlab 1200 Operator's Manual page 382

Table of Contents

Advertisement

E-22
Table E-18
Level Mode
1
2
3
4
5
6
1
2
3
4
5
6
*
WRSD = within-run standard deviation
Table E-19
Level Mode
1
2
3
4
5
6
7
8
9
1
02087462 Rev.V
Rapidlab 1245 Recovery and Precision Testing—pH
n
WRSD
pH Only
24
0.007
Capillary
pH Only
24
0.004
Capillary
pH Only
24
0.008
Capillary
pH Only
24
0.003
Capillary
pH Only
24
0.005
Capillary
pH Only
24
0.004
Capillary
pH Only
48
0.004
Syringe
pH Only
48
0.003
Syringe
pH Only
48
0.002
Syringe
pH Only
48
0.003
Syringe
pH Only
48
0.002
Syringe
pH Only
48
0.003
Syringe
Rapidlab 1245 Recovery and Precision Testing—pCO
n
WRSD
Capillary
24
0.16
Capillary
24
0.29
Capillary
24
0.56
Capillary
24
0.63
Capillary
24
0.94
Capillary
24
1.39
Capillary
23
3.58
Capillary
24
2.98
Capillary
24
3.27
Syringe
48
0.22
Rapidlab 1200 Operator's Guide: System Specifications
*
Observed
Expected
7.024
7.032
7.198
7.202
7.362
7.354
7.414
7.410
7.542
7.534
7.653
7.630
7.016
7.032
7.189
7.202
7.350
7.354
7.389
7.410
7.517
7.534
7.631
7.630
*
Observed
Expected
7.5
7.1
22.4
21.4
27.3
28.6
35.3
35.6
49.6
50.0
69.5
71.4
95.4
99.7
133.3
135.7
215.7
228.0
7.4
7.2
Recovery
WRCV
99.88
0.10
99.94
0.05
100.11
0.10
100.05
0.04
100.11
0.07
100.30
0.05
99.77
0.06
99.82
0.05
99.95
0.03
99.72
0.04
99.78
0.03
100.01
0.04
2
Recovery
WRCV
104.7
2.1
104.5
1.3
95.4
2.1
99.2
1.8
99.2
1.9
97.4
2.0
95.6
3.8
98.3
2.2
94.6
1.5
104.0
3.0