Siemens ARCADIS Avantic Maintenance Instruction
Hide thumbs Also See for ARCADIS Avantic:

Advertisement

ARCADIS Avantic
DHHS Maintenance Instructions
Print No.: SPR2-330.662.01.02.02
Replaces: SPR2-330.662.01.01.02
SP
© Siemens AG 2005
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model
_or_ design,_are_
reserved.
English
Doc. Gen. Date: 05.06

Advertisement

Table of Contents
loading

Summary of Contents for Siemens ARCADIS Avantic

  • Page 1 ARCADIS Avantic DHHS Maintenance Instructions © Siemens AG 2005 The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent...
  • Page 2 Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho- rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service procedures.
  • Page 3: Table Of Contents

    Tube Housing Assemblies ......6 - 2 Siemens AG SPR2-330.662.01 Page 3 of 4 ARCADIS Avantic Medical Solutions Rev. 02 05.06...
  • Page 4 7 ______ Final steps ____________________________________________________7 - 1 Final visual and function checks ......7 - 1 ARCADIS Avantic SPR2-330.662.01...
  • Page 5: Text Emphasis

    Symbols Checks and adjustments that must be performed with radiation ON are identified by the radiation warning symbol. Siemens AG SPR2-330.662.01 Page 1 of 4 ARCADIS Avantic Medical Solutions Rev. 02 05.06 CS PS SP...
  • Page 6: Safety Information And Protective Measures

    When performing the checks and maintenance work steps, please note: the product-specific safety information in this document, the safety information in the register Safety of the ARCADIS Operating Instructions. ARCADIS Avantic SPR2-330.662.01 Page 2 of 4 Siemens AG Rev. 02 05.06...
  • Page 7 This product is released for use in operating rooms and can be contaminated with infec- tious blood or other body secretions. Avoid all contact with blood or other body secretions! Siemens AG SPR2-330.662.01 Page 3 of 4 ARCADIS Avantic Medical Solutions Rev. 02 05.06 CS PS SP...
  • Page 8 1 - 4 Safety Information This page intentionally left blank. ARCADIS Avantic SPR2-330.662.01 Page 4 of 4 Siemens AG Rev. 02 05.06 CS PS SP Medical Solutions...
  • Page 9: After Completion Of Maintenance

    If the results of any of the tests described in this document exceed the tolerances given, the required adjustments must be performed. Contact Siemens service. After performing any adjustment, all applicable tests must be repeated. Customer: Customer No.:...
  • Page 10: Required Reference Documents

    (70cm distance from the focal spot). NOTICE This value is used because in typical applications the object to be examined is located approx. 30cm in front of the I.I. ARCADIS Avantic SPR2-330.662.01 Page 2 of 2 Siemens AG Rev.
  • Page 11: Radiation Protection

    Confirm that the certification labels as well as serial number and model number labels are legible. Use the illustrations which are shown in the Operating Instructions of the product. If labels are required, contact the SIEMENS Medical Solutions Local Service Organization or the SIEMENS Medical Solutions Regulatory Affairs department.
  • Page 12 3 - 2 Visual checks This page intentionally left blank. ARCADIS Avantic SPR2-330.662.01 Page 2 of 2 Siemens AG Rev. 02 05.06 CS PS SP Medical Solutions...
  • Page 13: Checking The "Deadman" Switch Function

    Check that the radiation indicators at the control console and on the monitor cart are visi- ble during radiation and that an audible (whistling) signal is sounded after exposure release for about 1.5 seconds. Radiation indicators are functioning Siemens AG SPR2-330.662.01 Page 1 of 2 ARCADIS Avantic Medical Solutions Rev. 02 05.06 CS PS SP...
  • Page 14: Fluoroscopy Timer

    Confirm that tube voltage and current are continuously indicated during fluoroscopy. Indicators functioning 5 Minimum source to skin distance Confirm that a spacer limiting the source to skin distance (SSD) to 30 cm is present. Spacer present ARCADIS Avantic SPR2-330.662.01 Page 2 of 2 Siemens AG Rev. 02 05.06...
  • Page 15: Checking The Accuracy Of The Tube Voltage

    The actual kV value must agree with the indicated value within the limits stated by & Siemens (max. deviation 10 % for the ARCADIS Avantic), if the X-ray system is con- nected to a power supply corresponding to Siemens specifications. Tools: PTW-Nomex...
  • Page 16: Calculations

    Divide the square root S by the average value x determined in step 1: Coefficient of variation C = ≤ C = ....(must be 0.05) ARCADIS Avantic SPR2-330.662.01 Page 2 of 6 Siemens AG Rev. 02 05.06...
  • Page 17: Entrance Dose Rate

    ” button at the operating console. Select 125 kV. Select full I.I. format, open the collimator to max. aperture, switch fluoros- copy ”ON”. Measure the dose and record. Siemens AG SPR2-330.662.01 Page 3 of 6 ARCADIS Avantic Medical Solutions Rev. 02 05.06 CS PS SP...
  • Page 18: Check The I.i. Format

    Attach a center cross with lead ruler to the center of the I.I. input screen. Switch on fluoroscopy. Read the size of the X iris on the monitor. Smallest field size at the I.I. input: ..cm x ..cm ARCADIS Avantic SPR2-330.662.01 Page 4 of 6 Siemens AG Rev.
  • Page 19: Checking The Field Limitation

    ∆ = deviation of A to B x = deviation in the x-axis y = deviation in the y-axis ∆ Fig. 4 Siemens AG SPR2-330.662.01 Page 5 of 6 ARCADIS Avantic Medical Solutions Rev. 02 05.06 CS PS SP...
  • Page 20 C-arm to the right must be < 47 mm (equal to > 80% over- ...... mm lapping of the visible area) ARCADIS Avantic SPR2-330.662.01 Page 6 of 6 Siemens AG Rev. 02 05.06 CS PS SP Medical Solutions...
  • Page 21: Technical Data For Dhhs Components

    15,2 mA maximum current Subtraction: 125 kV maximum voltage 15,2 mA maximum current Digital Cine Mode: 125 kV maximum voltage 250 mA maximum current Siemens AG SPR2-330.662.01 Page 1 of 2 ARCADIS Avantic Medical Solutions Rev. 02 05.06 CS PS SP...
  • Page 22: Maximum Deviations

    As measured in the mA-measuring pins D1.X39-D1.X40 (x97 opened) in the tube current circuit near ground with an mA-meter (FLUKE 8060 A or equivalent), minus the current flowing through the built-in high-voltage divider of 0.0025 mA/kV ARCADIS Avantic SPR2-330.662.01 Page 2 of 2 Siemens AG Rev.
  • Page 23 - there is no danger of collision for moving parts - safety distances are maintained - all parts are properly secured - there is no perceptible damage, etc. Siemens AG SPR2-330.662.01 Page 1 of 2 ARCADIS Avantic Medical Solutions Rev. 02 05.06 CS PS SP...
  • Page 24 7 - 2 Final steps This page intentionally left blank. ARCADIS Avantic SPR2-330.662.01 Page 2 of 2 Siemens AG Rev. 02 05.06 CS PS SP Medical Solutions...