Removing And Storing The Sensors; Removing Peripherals And Disinfecting The Exterior Surfaces; Shutting Down And Packing The System - Siemens Rapidlab 1200 Operator's Manual

Table of Contents

Advertisement

5-60

Removing and Storing the Sensors

1. Select
2. To start the procedure, select
3. Remove the sensor:
a. Lift the front cover, push up the latches on the measurement module door, and
b. Push the spring-loaded latch to the right.
c. Grasp the tab on the sensor and pull the sensor up and out.
4. Store each sensor in the original packaging.
Removing Peripherals and Disinfecting the Exterior
Surfaces
1. Remove the waste bottle and dispose of it according to your institution's protocol.
2. Clean and disinfect the exterior surfaces.
Refer to Cleaning and Disinfecting the Exterior Surfaces‚ page 5-2.
3. Remove the roll of printer paper and remove any remaining paper from the printer by
turning the paper-advance knob clockwise.

Shutting Down and Packing the System

1. Select
2. After the system displays the Shutdown screen, turn the power off and disconnect the
power cord from the electrical outlet and from the system.
3. If the optional barcode scanner is installed, disconnect the barcode scanner.
4. Pack the system in the original shipping carton.
If the original carton is no longer available, contact your local technical support
provider or distributor for a replacement shipping carton.
02087462 Rev. V
Status > Maintenance > Performing Sensor Maintenance
then lift the door.
Status > Shutdown > Yes
Rapidlab 1200 Operator's Guide: Maintenance
.
Continue
.
.

Advertisement

Table of Contents
loading

Table of Contents