Siemens Rapidlab 1200 Operator's Manual page 374

Table of Contents

Advertisement

E-14
Table E-11
Level Mode
1
2
3
4
5
6
1
2
3
4
5
6
*
WRSD = within-run standard deviation
Table E-12
Level Mode
1
2
3
4
5
6
7
8
9
1
02087462 Rev.V
Rapidlab 1240 Recovery and Precision Testing—pH
n
WRSD
pH Only
24
0.014
Capillary
pH Only
21
0.007
Capillary
pH Only
24
0.005
Capillary
pH Only
21
0.006
Capillary
pH Only
24
0.006
Capillary
pH Only
21
0.004
Capillary
pH Only
42
0.004
Syringe
pH Only
48
0.003
Syringe
pH Only
46
0.005
Syringe
pH Only
48
0.003
Syringe
pH Only
42
0.004
Syringe
pH Only
45
0.004
Syringe
Rapidlab 1240 Recovery and Precision Testing—pCO
n
WRSD
Capillary
24
0.17
Capillary
24
0.42
Capillary
24
0.8
Capillary
24
0.55
Capillary
24
0.78
Capillary
24
1.71
Capillary
24
1.25
Capillary
24
3.64
Capillary
24
6.27
Syringe
48
0.23
Rapidlab 1200 Operator's Guide: System Specifications
*
Observed
Expected
7.031
7.032
7.221
7.202
7.371
7.354
7.434
7.41
7.551
7.534
7.657
7.63
7.008
7.032
7.181
7.202
7.336
7.354
7.388
7.41
7.506
7.534
7.621
7.63
*
Observed
Expected
7.1
7.1
22.6
21.4
27.1
28.6
34.8
35.6
50.5
50
68.4
71.4
97
99.8
128.1
135.7
212.5
228
7
7.2
Recovery
WRCV
99.99
0.2
100.27
0.1
100.24
0.06
100.32
0.08
100.22
0.08
100.35
0.05
99.65
0.05
99.7
0.04
99.76
0.07
99.7
0.03
99.62
0.06
99.88
0.06
2
Recovery
WRCV
99.1
2.4
105.6
1.9
94.7
3
97.8
1.6
101.1
1.5
95.8
2.5
97.2
1.3
94.4
2.8
93.2
3
97.8
3.2