D50 Glucose Sensor Error - Siemens Rapidlab 1200 Operator's Manual

Table of Contents

Advertisement

Rapidlab 1200 Operator's Guide: Troubleshooting

D50 Glucose Sensor Error

Problem
The system detects
an open connection
in the glucose
biosensor.
Corrective Action
1.
Check that the biosensor is correctly installed, the contacts
are aligned, and the spring-loaded latch is closed.
2.
If the biosensor has been on the system less than one day,
perform three 2-point calibrations.
Refer to Performing Manual Calibrations‚ page 3-4.
3.
If the problem persists, replace the biosensor.
4.
If the problem persists, call for technical assistance.
6-69
02087462 Rev. V

Advertisement

Table of Contents
loading

Table of Contents