Checking The Level Of Fill Solution - Siemens Rapidlab 1200 Operator's Manual

Table of Contents

Advertisement

Rapidlab 1200 Operator's Guide: Maintenance
6. Aspirate the conditioner:
a. Draw the conditioner solution into a syringe and insert the syringe into the
b. Select
c. When prompted, remove the syringe and select
7. If you have a Rapidlab 1260 or 1265 system, replace the test/blank sensors
with the biosensors.
8. Select
9. Select
10. If you have a Rapidlab 1260 or 1265 system, select
11. If you have a Rapidlab 1240 or 1245 system, perform a 2-point calibration.
Refer to Performing Manual Calibrations‚ page 3-4.
12. Analyze a minimum of 2 levels of quality control material to verify sensor
performance.

Checking the Level of Fill Solution

Use this procedure to ensure that the sensors contain the correct level of fill
solution.
The pO
fill solution in the pO
The Glucose and Lactate biosensors do not require fill solution.
BIOHAZARD:
Refer to Appendix A, Protecting Yourself from Biohazards for recommended
precautions when working with biohazardous materials.
CAUTION:
without first discharging static buildup. Touch the inner surface of the
measurement module frame to discharge static buildup.
1. Select
2. Select
3. Lift the door to the measurement module.
Refer to Preparing for Maintenance Procedures‚ page 5-1.
sample port.
.
Continue
The conditioning cycle runs for 10 minutes.
.
Return
.
Analyze
and pCO
sensors do not require maintenance. Slight discoloration of the
2
2
and pCO
2
Wear personal protective equipment. Use universal precautions.
Do not remove or return the sensors to the measurement module
Status > Maintenance > Performing Sensor Maintenance
.
Continue
sensors is normal.
2
.
Continue
.
Yes
.
02087462 Rev. V
5-7