Siemens Rapidlab 1200 Operator's Manual page 67

Table of Contents

Advertisement

Rapidlab 1200 Operator's Guide: System Overview and Intended Use
The measuring electrode contains platinum and glucose oxidase or lactate oxidase
in a binder, while the reference electrode is composed of Ag/AgCl. Two other
electrodes are also present. The counter electrode is a Pt (platinum) conductor that
ensures a constant applied potential. Another measuring electrode, without the
enzyme, determines interfering substances in the sample. The potential from
interfering substances is removed from the total differential measurement. A
microporous cover membrane separates the electrodes from the sample.
1
2
3
4
5
Figure 1-35
A constant voltage, called a polarizing voltage, is maintained during analysis. In
the glucose sensor, glucose from the sample interacts with the glucose oxidase on
the surface of the measuring electrode to form hydrogen peroxide and gluconic
acid:
where GOX is the glucose oxidase.
The polarizing voltage is sufficient to cause oxidation of the hydrogen peroxide to
oxygen:
Glucose or lactate measuring electrode
Reference electrode
Interference measuring electrode
Counter electrode
Sample path
Glucose and Lactate Biosensor
+
+
C
H
O
H
O
O
6
12
6
2
2
(
glucose
)
GOX
+
C
H
O
H
O
6
12
7
2
2
(
gluconic
acid
)
1-47
02087462 Rev. V

Advertisement

Table of Contents
loading

Table of Contents