Potentiometers and Reference Switches

Cone drive TR, reference switch S 9 and potentiometer r9

NOTE
Connect the oscilloscope to D5.X2.a23 (for S9) (0 V to X30); 5 V/div, 0.1 s/div.
Switch the system on.
Move the shock wave head into the therapy position.
Check the reference voltage, if necessary set it with R220 on D8. (U
mV; check on D5 MP X31, 0V at X30)
NOTICE
Connect a DVM to D5.X2.a24 (0 V at X30).
Check the function of potentiometer r9.
Prerequisite: US probe not inserted.
Move to the reference position in Normal mode using the tableside control panel.
Use the "Cone rotation right or left" key for this.
When reference switch S9 is actuated, the DVM should display a reading of 2.560 V ± 2
mV. ADC value = 512
Check the switching hysteresis of the reference switch S9 according to
Fig. 8:

Setting the parameters

Refer to the "Parameter Groups" section of the "Service software, Software parameter /
Monitoring" instructions RXL2-120.032.02..
Siemens AG
Medical Solutions
Only perform if the ultrasound option is available.
Changing the reference voltage
When changing the reference voltage D5 U
5 mV, check the switching hysteresis for all drives.
The cone rotation drive travels to the 0° position and stops.
The 0° marks on the toothed ring and the lifting cylinder line up.
(If in "unregulated motion" mode, readjust the cone rotation )
The position is displayed on the US tableside control panel.
Switching hysteresis S9
RXL2-120.841.02.01.02
09.05
ref
Page 43 of 72
CS PS 24
43
0
= 5.120 V ± 5
= 5.120 V ±
ref
(Fig. 8 / p.
43).
0