Rapidlab 1200 Operator's Guide: Troubleshooting

Using the CO-ox Screen

Only Rapidlab 1245 and 1265 systems use the functions on the CO-ox screen.
The Diagnostics–CO-ox screen displays buttons for performing the following
actions:
testing the lamp calibration
testing wavelength calibration
testing the CO-ox Sample Chamber
turns the CO-ox lamp off or on

Performing the Lamp Calibration Test

This test is only performed on Rapidlab 1245 and 1265 Systems.
Perform this test whenever you change the CO-ox lamp or whenever you remove
and reinstall the lamp. You can also use this test to perform a lamp calibration.
The system displays the pass or fail results on a printed report.
1. Select
2. Enter your password, if necessary.
3. Select
4. Wait while the system performs the test.
The system prints test results as Passed or Failed.
5. Select
6. To return to the Status screen, select
The system performs a full calibration.
.
Status > Diagnostics
.
CO-ox > Lamp Cal
.
Return
.
Return
6-45
02087462 Rev. V