Rapidlab 1200 Operator's Guide: Maintenance
b. Push the spring-loaded latch to the right, grasp the tab on the reference
1
2
Figure 5-11
c. Align the top of the test/blank reference sensor (TB5) with the sensor
4. If you have a Rapidlab 1260 or 1265 system, replace the Glucose and Lactate
biosensors with the test/blank Glucose Lactate (TB4) sensors.
Refer to step 3.
5. Close the measurement module door and then close the front cover.
6. Aspirate the deproteinizer:
a. Invert the deproteinizer vial several times to mix the deproteinizer
b. Draw the deproteinizer into a syringe and insert the syringe into the
c. Select
d. When prompted, remove the syringe and select
sensor, and pull the reference sensor up and out.
Reference sensor
Spring-loaded latch
Measurement Module
contact and snap the sensor into place.
solution.
sample port.
.
Continue
The system displays a message prompting you to condition the sensors.
5-23
.
Continue
02087462 Rev. V