Rapidlab 1200 Operator's Guide: Troubleshooting

D35 Codes

D35 Electronics Error: 1-3, 7-12

Problem
An error has
occurred in the
electronic
components.

D35 Electronics Error: 4

Problem
An error has
occurred in the
electronic
components of the
heater.

D35 Electronics Error: 13

Problem
An error has
occurred in the
electronic
components of the
pump motors.

D36 Codes

D36 Cartridge Loading Error

Problem
The cartridge
loader fails to move
to the correct
position during
reagent or wash
cartridge removal
or installation.
Corrective Action
1.
Select
Status > Shutdown.
2.
Wait 10 seconds, and then turn the system on.
3.
If the problem persists, call for technical assistance.
Corrective Action
1.
Ensure that the fans are working and nothing is blocking the
flow of air.
2.
Inspect the air filters and replace them if required.
3.
Select
Status > Shutdown
4.
Wait 10 seconds, and then turn the system on.
5.
Select
Diagnostics > Voltage and Temperatures
6.
If the problem persists, call for technical assistance.
Corrective Action
1.
Select
Diagnostics > Pumps and Valves > Sample Pump
2.
Select
Status > Shutdown
3.
Wait 10 seconds, and then turn the system on.
4.
If the problem persists, call for technical assistance.
Corrective Action
1.
Ensure that the cartridge is correctly positioned in the door
and nothing is blocking the cartridge from loading.
Use the Eject Reagent Cartridge or Eject Wash
NOTE:
Cartridge test depending on the affected cartridge.
2.
Select
Diagnostics > Cartridge > Eject W Cartridge
.
Eject R Cartridge
3.
If the problem persists, replace the cartridge.
4.
If the problem persists, call for technical assistance.
.
.
.
02087462 Rev. V
6-65
.