6-20
b. If using a standard syringe, hold the syringe firmly.
CAUTION:
with great force through the port at the back of the sample connector.
5. Press the syringe plunger gently while holding gauze against the port at the back of the
sample connector (item 1, Figure 6-7), to catch the obstruction and any excess liquid.
6. Remove the syringe.
BIOHAZARD:
practices of your workplace.
7. Discard the gauze.
1
2
3
4
5
Figure 6-7
8. Reinstall the reagent manifold.
a. Insert the reagent manifold until it latches into place, using the 2 latches (item 5,
02087462 Rev. V
If the path is totally obstructed, liquid may be expelled suddenly and
Discard the gauze according to the established biohazard disposal
Fitting to which the reagent manifold tubing and CO-ox pump tubing are connected
(Rapidlab 1245 and 1265 systems)
Reagent manifold tubing that connects the fitting to the port on the top, left-side of the
reagent manifold (Rapidlab 1245 and 1265 systems)
Sample connector
Waste tubing connected to port on top, right-side of the reagent manifold
Two latches that secure the reagent manifold to the system
Reagent Manifold and Tubing Connections
Figure 6-7) that secure the manifold to the system.
Rapidlab 1200 Operator's Guide: Troubleshooting