Calibration Procedure Using Agilent 8480 Series Power Sensors - Agilent Technologies E4419B User Manual

Dual channel power meter
Hide thumbs Also See for E4419B:
Table of Contents

Advertisement

Note
Example
To calibrate both channels of the power meter with Agilent E-series power
sensors connected to both channels.
Calibration Procedure using Agilent 8480 Series Power
Sensors
The following procedure describes how you calibrate the power meter with
the Agilent 8480 series power sensors.
Note
Agilent E4419B User's Guide
The following procedure details the calibration for channel A. To
calibrate channel B use the same procedure using the equivalent
channel B softkeys.
Zero
1. Press
.
Cal
2. Connect the power sensor to the POWER REF output.
3. Press
Cal Cal A
,
calibration the wait symbol is displayed. (The power meter
automatically turns on the POWER REF output.)
Zero
Press
,
Cal
Cal
Connect the channel A power sensor to the POWER REF output.
Press
then wait till the calibration is completed.
Cal A
Remove the channel A power sensor from the POWER REF
output.
Connect the channel B power sensor to the POWER REF output.
Press
then wait till the calibration is completed.
Cal B
V8486A and W8486A sensors
For most 8480 series sensors the correct (A type or D type)
linearity correction table is automatically selected. However, for
the V8486A and W8486A sensors the automatic selection must be
overriden and the D type correction selected. Subsequent
connection of another A type sensor will result in a warning
message stating that "Linearity Override May be Required".
To select the linearity type to be applied:
Zeroing and Calibrating the Power Meter
to calibrate the power meter. During
.
Power Meter Operation
2-9