:MARKer:MODE

Command Syntax
:MARKer:MODE <mode>
<mode> ::= {OFF | MEASurement | MANual}
The :MARKer:MODE command sets the cursors mode. OFF removes the
cursor information from the display. MANual mode enables manual
placement of the X and Y cursors. In MEASurement mode the cursors
track the most recent measurement.
If the front- panel cursors are off, or are set to the front- panel Hex or
Binary mode, setting :MARKer:MODE MANual will put the cursors in the
front- panel Normal mode.
Setting the mode to MEASurement sets the marker sources
(:MARKer:X1Y1source and :MARKer:X2Y2source) to the measurement
source (:MEASure:SOURce). Setting the measurement source remotely
always sets the marker sources.
Query Syntax
:MARKer:MODE?
The :MARKer:MODE? query returns the current cursors mode.
Return Format
<mode><NL>
<mode> ::= {OFF | MEAS | MAN}
See Also
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
(see
page
586)
"Introduction to :MARKer Commands"
":MARKer:X1Y1source"
":MARKer:X2Y2source"
":MEASure:SOURce"
on page 242
on page 245
on page 247
on page 279
Commands by Subsystem