Adjusting the sprayer tip position:
Y control
Y scale
3.
While viewing the camera control, adjust the X, Y, and Z thumbscrew
controls on the adjuster assembly to move the sprayer tip relative to the
sampling cone and baffle, and adjust for optimal signal.
Sprayer tip control parameters:
Axis
X
Y
Z
4.
Mark the optimal position on the screen using the cross-hairs on the
viewing screen.
The marked emitter position provides a guide when you insert the
Tip:
analytical nanoTile, which should be within 0.25 mm.
2-16
Preparing the System
Direction
Left () to right (+)
Up (+) and down ()
Towards (+) or away from () sample cone
X control
X scale
Z scale
Z control
Range
3 to +2 mm
+6 to 6 mm
+2 to 11 mm