4 Maintenance Procedures
13. Use the small Phillips screwdriver to fit and tighten the repeller
electrode retaining screw.
14. Insert the UV bulb into the lamp drive assembly and push it fully home.
15. Fully retract the lamp mounting shaft from the source enclosure.
16. Refit the lamp-assembly collar-cover, and secure it on its base with the 4
screws.
17. Refit the lamp assembly cover, and secure it on its base (2 screws) and
above the bulb extraction aperture (2 screws).
18. Refit the source enclosure release handle, and secure it with the 2
screws.
19. Refit and reconnect the source enclosure to the machine, refit the corona
pin and probe. Refer to
page 72

Replacing the ionKey source fluid lines

If the inlet, outlet, or waste lines become damaged, replace them, to ensure
optimal performance. Each fluid line is supplied with a pre-fitted PEEK™
connector assembly at the point where it connects to the ionKey source
module. You must replace the capillary tubing and the PEEK assembly as a
single unit. No parts of the fluid line are reusable once detached.