8.
Disconnect the high-voltage cable (white) from the high-voltage supply
outlet on the mass spectrometer.
9.
Disconnect the reference probe power cable (green) from the reference
probe power inlet on the mass spectrometer.
10. Disconnect the fluid infusion line from the onboard IntelliStart Fluidics
on the mass spectrometer.
11. Carefully lift off the TRIZAIC UPLC source module and store safely.
12. For a SYNAPT mass spectrometer, remove the TRIZAIC source adaptor
housing's screws, and then remove the housing.
3-10
Maintenance Procedures