Removing an analytical nanoTile

To remove an analytical nanoTile:
1.
Stop the solvent flow using one of these methods.
Press the Stop Flow button located behind the TRIZAIC source door.
In the console, select Sample Manager > Maintain > Remove
nanoTile.
The pressure LED blinks green as the system is decompressing.
Requirement:
2.
After the pressure LED goes dark, rotate the lever clockwise 180 degrees
to the vertical position, with the unlock indicator in the 12 o'clock
position.
3.
Gently pull the nanoTile from the docking port.
4.
Place the nanoTile in the protective sheath.
Do not proceed until the pressure LED is off.