Waters TRIZAIC UPLC System Manual page 19

Table of Contents

Advertisement

To remove the source enclosure:
Warning:
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
To avoid electric shock, prepare the instrument for work
Warning:
performed on its source before commencing this procedure.
1.
Prepare the instrument for working on its source (see
Warning:
take great care while working with these components.
Warning:
an ESI probe; the probe tip is sharp.
2.
Slide the instrument's source interface door open.
3.
Disconnect the probe options cable from the instrument's connectors.
Warning:
working with the source enclosure open when a corona pin is
fitted; the corona pin tip is sharp.
Caution:
enclosure, to avoid damaging the sample inlet, you must move the
sprayer platform out of the source enclosure before you open the
source enclosure.
4.
Pull the source enclosure release (located at the bottom, right-hand side)
outward, and swing open the enclosure.
5.
Grasp the source enclosure using two hands, and lift it vertically off the
two supporting studs on the source adaptor housing (see
"Removing the source enclosure:" on page
The source components can be contaminated with
The probe and source can be hot. To avoid burn injuries,
To avoid puncture wounds, take great care if removing
To avoid puncture wounds, take great care while
If you are removing a NanoLockSpray™ source
Installing the TRIZAIC UPLC source
page
2-1).
figure
2-4).
2-3