slightly using the Z control, or move the emitter away from the baffle
under the X control.
Arcing from the sprayer tip to the baffle:
Requirement:
good spray.
Tips:
Fine tune the position of the sprayer while acquiring a spectrum of a
standard compound. Small adjustments to the sprayer position can
make large differences to the source sensitivity.
If spray instability occurs, you can clean the emitter by washing
with isopropanol and blowing it dry with air.
6.
Close the TRIZAIC source door.
7.
In MassLynx software, create a shutdown method that puts the
instrument in Standby mode when an LC error terminates the flow.
Result:
the emitter in the absence of flow.
See also:
shutdown method.
2-18
Preparing the System
Arcing
The capillary voltage must be high enough to maintain a
The shutdown method turns the capillary voltage off, to preserve
MassLynx online Help for information about creating a