TRIZAIC UPLC source:

nanoTile

docking port
The dual electrospray ion source enables the optimized co-introduction of
sample and lock-mass compound directly into the ion source. This feature
provides authenticated, exact-mass measurement in both MS and MS/MS
modes at low flow rates.
The TRIZAIC source holds a nanospray probe positioned orthogonally to the
nanoTile. A motorized baffle rotates to admit spray to the sampling cone.
Spray indexing permits acquiring sample and lock-spray data in separate
data channels. The baffle design ensures negligible cross-talk between the two
sprays. Data from the lock spray are used to calculate a correction factor for
the mass-scale calibration, which is then applied to the sample data, to
provide exact-mass information.
The sprayer platform adjuster assembly allows you to precisely position the
emitter in X-, Y-, and Z-dimensions.
nanoTile
The nanoTile eliminates multiple fluid connections and minimizes the
potential for extra-column dispersion.
Each nanoTile is a high-pressure, microfluidic device capable of withstanding
UPLC pressures. Produced using a modified, co-fired ceramic manufacturing
process, it contains an integrated electrospray emitter and packed, analytical
and trapping channels.
1-4
TRIZAIC UPLC System Capabilities
Camera mount
nanoTile
locking/unlocking
handle
Front cover