Installing Adaptor Housing On A Synapt Source - Waters TRIZAIC UPLC System Manual

Table of Contents

Advertisement

Removing the source enclosure:
Cable
storage
positions
Source
enclosure
release
6.
Store the cables neatly by plugging them into the cable storage positions
on the rear of the source enclosure.
7.
Perform one of these tasks, depending on whether you must install the
source adaptor housing:
Install the source adaptor housing preliminary to installing the
source (SYNAPT).
Continue with

Installing adaptor housing on a SYNAPT source:

After removing the SYNAPT source enclosure, install the SYNAPT source
adaptor housing, which provides supporting studs for the TRIZAIC source.
To install source adaptor housing on a SYNAPT mass spectrometer:
1.
Unscrew the front plate from the source enclosure.
2.
Remove the SYNAPT source.
3.
Install the source adaptor housing, which provides two supporting studs
for the TRIZAIC source, and tighten the screws.
2-4
Preparing the System
"Installing the TRIZAIC UPLC source" on page
Supporting studs
Source
adaptor
housing
2-5.

Advertisement

Table of Contents
loading