Installing A Fluid Line - Waters TRIZAIC UPLC System Manual

Table of Contents

Advertisement

4.
Cut off and discard the capillary tube fitting at the UPLC-instrument
end of the fluid line.
5.
Return to the TRIZAIC fluid line connector plate, and gently pull off the
O-ring from around the fluid line compression fitting.
6.
Grasping the PEEK fluid connector, draw the fluid line out and away
from the connector plate.
7.
Dispose of all parts of the fluid line assembly in accordance with
standard laboratory procedures for contaminated vessels and sharps.

Installing a fluid line

Required materials
Chemical-resistant, powder-free gloves
Fluid line assembly
To install a fluid line:
Warning:
performed on its source before commencing this procedure.
1.
Prepare the instrument for working on its source (see
2.
Pull open the source enclosure door, and then locate the fluid line
connector plate.
3.
Identify the correct hole in the connector plate and thread the loose
(unfitted) end of the fluid line through the aperture until the O-ring and
compression fitting protrude from the left-hand side of the fluid
connector plate (see
3-6
Maintenance Procedures
The probe and source components can be contaminated
Warning:
with biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
To avoid electric shock, prepare the instrument for work
The probe and source can be hot. To avoid burn injuries,
Warning:
take great care while working with these components.
figure "Fluid line connections:" on page
page
2-1).
3-4).