To avoid electric shock,
Warning:
• do not remove the instrument's panels. There are no user-serviceable
items inside the instrument.
• ensure that the instrument is in Standby mode before commencing
any maintenance.
The probe and source can be hot. To avoid burn injuries, take
Warning:
great care while working with these components.
When performing maintenance inside the source enclosure,
Caution:
ensure that the following criteria are met:
• Instrument is in Standby mode.
• LC flow is diverted to waste or set to off.
• Desolvation gas is turned off.
To prevent damage to the nanoTile,
Caution:
• handle it with care. The components are fragile.
• do not subject it to pressures that exceed 34,474 kPa (345 bar,
5000 psi).
• do not apply electrospray potential to the emitter without flow.
• do not drop it, or immerse it in liquid.
• do not freeze or overheat it. Keep the nanoTile within the allowed
temperature ranges during operation and in storage.
• use the nanoTile sheath to protect the device when it is not in use.
• do not bend or pull the capillary connection tubing at the TRIZAIC
module coupling.
• avoid excess voltage, which can erode the emitter over time.
• do not touch the electrospray emitter, which can bend.
See
Appendix A
for safety advisory information.
nanoACQUITY UPLC System Operator's Guide and the
See also:
documentation accompanying the mass spectrometer for detailed information
about safety advisories.
Safety and handling
3-3