The nanoACQUITY UPLC sample manager powers the TRIZAIC
Tip:
UPLC source.
3.
Power-on the solvent managers.
Result:
series of startup tests.
4.
Ensure the mass spectrometer is powered-on.
5.
Monitor the system instrument power LEDs and status LEDs.
Full initialization can require as long as 7 minutes.
Tips:
Status LEDs
Status LEDs serve in these applications:
Flow LED on solvent managers
Operate LED on mass spectrometers
Status and Pressure LEDs on the TRIZAIC source
LEDs on the TRIZAIC source:
TRIZAIC LED colors and their meaning:
LED color
Unlit
Flashing
green
Constant
green
The Stop Flow button on the TRIZAIC UPLC source stops fluid flow and
depressurizes a nanoTile installed in the docking port. When the
2-10
Preparing the System
Each device or instrument beeps three times and then runs a
Status LED
The nanoTile is not
installed.
The nanoTile is in place,
and the system is waiting
for it to become operable.
The nanoTile is correctly
inserted in the source, and
the clamp is engaged.
Pressure LED
The TRIZAIC UPLC source
is removed or there is no
flow.
The source is pressurizing
or depressurizing.
The source is at operating
pressure.

Advertisement