17
Waveform Generator

To add noise to the waveform generator output

1 If the Waveform Generator Menu is not currently displayed on the
2 In the Waveform Generator Menu, press the Settings softkey.
3 In the Waveform Generator Settings Menu, press the Add Noise softkey and turn
Note that adding noise affects edge triggering on the waveform generator source
(see
output signal (which can be sent to TRIG OUT, see
OUT Source"
the noise source.

To add modulation to the waveform generator output

Modulation is where an original carrier signal is modified according to the
amplitude of a second modulating signal. The modulation type (AM, FM, or FSK)
specifies how the carrier signal is modified.
To enable and set up modulation for the waveform generator output:
1 If the Waveform Generator Menu is not currently displayed on the
2 In the Waveform Generator Menu, press the Settings softkey.
3 In the Waveform Generator Settings Menu, press the Modulation softkey.
4 In the Waveform Generator Modulation Menu:
250
oscilloscope's softkeys, press the [Wave Gen] key.
the Entry knob to select the amount of white noise to add to the waveform
generator output.
"Edge Trigger"
on page 140) as well as the waveform generator sync pulse
on page 284). This is because the trigger comparator is located after
oscilloscope's softkeys, press the [Wave Gen] key.
Press the Modulation softkey to enable or disable modulated waveform
generator output.
You can enable modulation for all waveform generator function types except
pulse, DC, and noise.
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
"Setting the Rear Panel TRIG