Navigating Segments - Keysight Technologies InfiniiVision 2000 X Series User Manual

Table of Contents

Advertisement

12
Acquisition Control
The oscilloscope runs and fills a memory segment for each trigger event. When
the oscilloscope is busy acquiring multiple segments, the progress is displayed in
the upper right area of the display. The oscilloscope continues to trigger until
memory is filled, then the oscilloscope stops.
If the signal you are measuring has more than about 1 s of inactivity, consider
selecting the Normal trigger mode to prevent AutoTriggering. See
Auto or Normal trigger mode"
See Also

Navigating Segments

1 Press the Current Seg softkey and turn the Entry knob to display the desired
You can also navigate segments using the [Navigate] key and controls. See
navigate segments"
190
Progress indicator
"Navigating Segments"
"Infinite Persistence with Segmented Memory"
"Segmented Memory Re-Arm Time"
"Saving Data from Segmented Memory"
segment along with a time tag indicating the time from the first trigger event.
on page 58.
on page 170.
Sample rate
on page 190
on page 191
on page 191
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
"To select the
on page 191
"To