CAN Serial Decode

To set up the oscilloscope to capture CAN signals, see
page 327.
For CAN triggering set up see
NOTE
To set up CAN serial decode:
1 Press [Serial] to display the Serial Decode Menu.
2 If the decode line does not appear on the display, press the [Serial] key to turn it
3 If the oscilloscope is stopped, press the [Run/Stop] key to acquire and decode
If the setup does not produce a stable trigger, the CAN signal may be slow enough that the
NOTE
oscilloscope is AutoTriggering. Press the [Mode/Coupling] key, then press the Mode softkey
to set the trigger mode from Auto to Normal.
You can use the horizontal Zoom window for easier navigation of the decoded data.