Verifying System Operation

System Set Up

CAUTION

Waveform Generation

NOTE
Prior to verifying system operation, the N8241A software must be installed on the
PC and the LAN line connected to the PC and AWG module. For more information
refer to
"Installing the Software" on page 25
.
LAN" on page 26
1. Connect the power cord to the AWG module and turn the power on.
Before switching on this instrument, make sure the supply voltage is in the specified
range.
An Agilent E4440A Spectrum Analyzer or equivalent is required to view the
waveforms.
1.
Connect the channel 1 positive (+) output to the spectrum analyzer RF input
connector.
2.
Open the user interface by double-clicking the N8241A Control Utility icon
placed on the desktop during installation.
3.
In the Output tab, configure the signal conditioning path to include the 500MHz
reconstruction filter through CH1 OUT (toggle the switches you want to
connect) on channel 1 and CH2 OUT on channel 2. The connection will
enable differential mode.
automatically
4. Click on the negative (-) node to enable single-ended mode. Notice that the
Chapter 1
Introducing the N8241/2A AWGs
Verifying System Operation
and
"Connecting to the AWG over the
29