This measurement is available for analog channels.

Rising Edge Count

The Rising Edge Count measurement is an edge count for the selected waveform
source.
This measurement is available for analog channels.

Falling Edges Count

The Falling Edges Count measurement is an edge count for the selected waveform
source.
This measurement is available for analog channels.

Measurement Thresholds

Setting measurement thresholds defines the vertical levels where measurements
will be taken on an analog channel or math waveform.
Changing default thresholds may change measurement results
NOTE
The default lower, middle, and upper threshold values are 10%, 50%, and 90% of the value
between Top and Base. Changing these threshold definitions from the default values may
change the returned measurement results for Average, Delay, Duty Cycle, Fall Time,
Frequency, Overshoot, Period, Phase, Preshoot, Rise Time, +Width, and -Width.
1 From the Measurement Menu, press the Settings softkey; then, press the
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
Negative pulse count
Thresholds softkey to set analog channel measurement thresholds.