Setup And Hold Trigger - Keysight Technologies InfiniiVision 2000 X Series User Manual

Table of Contents

Advertisement

10
Triggers
7 Press the Qualifier softkey to toggle between "less than", "greater than", or
8 If you selected the "less than" or "greater than" Qualifier, press the Time softkey;

Setup and Hold Trigger

The Setup and Hold trigger looks for setup and hold violations.
One oscilloscope channel probes the clock signal and another channel probes the
data signal.
To trigger on setup & hold violations:
1 Press the [Trigger] key.
2 In the Trigger Menu, press the Trigger softkey; then, turn the Entry knob to
3 Press the Clock softkey; then, turn the Entry knob to select the input channel
4 Set the appropriate trigger tevel for the clock signal using the Trigger Level
5 Press the Rising Edge or Falling Edge softkey to specify the clock edge being used.
6 Press the Data softkey; then, turn the Entry knob to select the input channel
154
None.
This lets you specify that a runt pulse be less than or greater than a certain
width.
then, turn the Entry knob to select the time.
Data
Clock
(rising
edge)
select Setup and Hold.
with the clock signal.
knob.
with the data signal.
Setup time
Hold time
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents