Reference Waveforms
4 Press the Save to R1/R2 softkey to save the waveform to the reference waveform
Reference waveforms are non-volatile — they remain after power cycling or performing a
NOTE
default setup.
To clear a
1 Press the [Ref] key to turn on reference waveforms.
reference
2 In the Reference Waveform Menu, press the Ref softkey and turn the Entry knob
waveform location
3 Press the Clear R1/R2 softkey to clear the reference waveform location.
Reference waveforms are also cleared by a Factory Default or Secure Erase (see
Chapter

To display a reference waveform

1 Press the [Ref] key to turn on reference waveforms.
2 In the Reference Waveform Menu, press the Ref softkey and turn the Entry knob
3 Then, press the Ref softkey again to enable/disable the reference waveform
98
location.
to select the desired reference waveform location.
18, "Save/Recall (Setups, Screens, Data)," starting on page 257).
to select the desired reference waveform location.
display.
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide