Vertical Controls

To specify the probe skew

When measuring time intervals in the nanoseconds (ns) range, small differences in
cable length can affect the measurement. Use Skew to remove cable-delay errors
between any two channels.
1 Probe the same point with both probes.
2 Press one of the probes associated channel key.
3 In the Channel Menu, press Probe.
4 In the Channel Probe Menu, press Skew; then, select the desired skew value.
Each analog channel can be adjusted ±100 ns in 10 ps increments for a total of
200 ns difference.
The skew setting is not affected by pressing [Default Setup] or [Auto Scale].
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
67